Equipment
··Class II medical device manufactured in Merge’s FDA-regulated, ISO 13485 certified
facility
Accuracy
··Complies with AAMI (Association for the Advancement of Medical Instrumentation)
Standards +/- 5 mmHg (millimeters of mercury)
Rate of Cuff Inflation
··Approximately 8 mmHg/sec
Temperature
··Operating: 50 to 104° F (10 to 40° C)
··Storage: -40 to 131° F (-40 to 55° C)
Humidity
··Operating: 10 to 95% (noncondensing)
··Storage: 10 to 95% (noncondensing)
Weight Measurement
··Range: 40-440 lbs.
··Accuracy: +/- 1 lbs.
Measurement Range
··Systolic: +50 to +260 mmHg
··Diastolic: +30 to +240 mmHg
··Pulse: 40 bpm - 199 bpm
Measurement Method
··Oscillometric
Overpressure Limit
··310 +/- 10 mmHg
Test Time
··Typically less than one minute
Physical Dimensions
··Height: 68.0 in (170.0 cm)
··Width: 33.5 in. (85 cm)
··Depth: 25.0 (63.5 cm)
Weight
··Model LC-500: 175 lbs (79.3 kg) (standard cabinet)
··200 lbs (90.7 kg) (wheelchair accessible cabinet)
Input Power
··117 VAC, 60 Hz, 2.5 Amps
··Power Supply Switch Selectable: 120/240 VAC, 50-60Hz
··3-wire grounded outlet
Weight Measurement
··Range: 40 to 440 lbs.
··Accuracy: +/- 1 lb.
Blood Oxygen Measurement
··Method: Infrared
··Accuracy: +/- 2%
Special Options
··Blood oxygen saturation, 2.25” printer, Barcode reader, Magnetic stripe reader, Wheelchair accessibility.
Data ports allow for downloading glucose meters, pedometers, etc.