Publicom, Inc.
Publicom, Inc.

Accuracy
Complies with Association for the Advancement of Medical
Instrumentation(AAMI) Standard SP10-1987


Rate of Cuff Inflation
Approximately 8 mmHg/sec


Temperature
Operating: 50 to 104° F (10 to 40° C)
Storage: -40 to 131° F (-40 to 55° C)


Humidity
Operating: 10 to 95% (non-condensing)
Storage: 10 to 95% (non-condensing)


Measurement Range
Systolic: +50 to +260 mmHg
Diastolic: +30 to +240 mmHg
Pulse: 40 bpm to 199 bpm


Measurement Method
Oscillometry


Overpressure Limit
310 +/- 10 mmHg


Test Time
Typically less than one minute


External meter tracks usage

 

Input Power
117 volts AC, 60 Hz, 2.5 amps
3-wire grounded outlet

 

Physical Dimensions

Height of 90550-01: 47.8 in (121.4 cm)

Height of 90550-15: 64.3 in (163.2 cm)

Width: 24.2 in (61.5 cm)
Depth: 20.3 in (51.6 cm)

Weight of 90550-01: 82 lbs (37.3 kg)
Weight of 90550-15: 90 lbs (40.1 kg)

 

Digital Displays
Systolic & diastolic pressure, pulse rate

 

Colors & Graphics
Multiple cabinet finishes (based on availability)
Custom graphics available. Contact us for more details on custom graphic packages.

 

Display Panel Inserts
4 – 7.5 in x 11.5 in